The GTC-1 is a specialized measurement device used to determine the gel time of various thermosetting resins and reaction polymers during the curing process. It utilizes a standardized plunging rod motion to detect the exact moment of viscosity change, providing essential data for laboratory research and industrial quality control.

  • Precise electronic timing for curing analysis

  • Standardized testing for repeatable batch results

  • Simplified operation for routine laboratory use

[acf_all_fields_accordion]

− Features

  • With memory for 50 tests
  • Max 24 Hour per measuring test
  • Data files in CSV format
  • Data exchange via USB stick
  • Illustrated spare parts list
  • Connections retro compatible (compatible with earlier versions)
  • Compact design, easy to operate
  • Reasonable price / performance ratio

+ Technical Specifications

+ Standards

Features
  • With memory for 50 tests
  • Max 24 Hour per measuring test
  • Data files in CSV format
  • Data exchange via USB stick
  • Illustrated spare parts list
  • Connections retro compatible (compatible with earlier versions)
  • Compact design, easy to operate
  • Reasonable price / performance ratio

Electrical Specifications

Power supply 115 VAC / 230 VAC, 50/60 Hz (internally switchable)
Power consumption 20 VA, 650 VA with heating
Connector 7 pole, N, P, PE and Pt100 2 wires, plug type Binder
Main switch on the rear panel
Fuses 2 pieces, F 3.15 A L 250 VAC
Electrical conformity EMV 2014/30/EU, EN 61010-1:2010, EN 61010-2-010:2014, EN 61326-1:2013

Temperature Control for the Electrical Heat Block GT

Actual value (Pt100) Range: 0.0 °C … 250.0 °C, adjustable and calibrated
Accuracy ± 0.3 % of full scale ± 1 digit
Set value Range: 0.0 °C … 200.0 °C
Resolution 0.1 K
Load output Power max. – Resistive load: maximum 600 W
self-optimization

Temperature measurement for the sample

Sensor Thermocouple Type K (Ni-CrNi), EN 60 584
Connector Mini connector type K thermocouple, EN 60 584
Range 0.0 °C 400.0 °C …
Accuracy ± 1.5 °C including cold-junction compensation
Adjustable and calibratable
Time Uncertainty up to 10 hours, ± 2 seconds
Up to 24 hours, ± 3 seconds

Test Unit with optical Sensor

Motor 24 VAC, 50/60 Hz, 6 rpm
Switch optical sensor, standard!
Sample temperature Thermocouple Type K for installation in the test tube to the sample
Stroke 25mm, 10 seconds for an up and down motion