Muffle Furnace upto 1100 °C

Ceramic Muffle For Aggressive Substances

Ceramic Muffle For Aggressive Substances

Ceramic Muffle For Aggressive Substances

Model No. :

Brand : Nabertherm

These compact muffle furnaces utilize heating elements integrated within ceramic muffles to provide exceptional resistance against aggressive vapors and chemical contaminants. The specialized design ensures superior temperature uniformity and durability for demanding laboratory applications requiring heat treatment up to 1100 °C.

  • Four sided heating for thermal uniformity

  • Embedded elements resist aggressive chemical vapors

  • Dual shell housing keeps surfaces cool

[acf_all_fields_accordion]

− Features

  • Tmax 1100 °C
  • Muffle heated from four sides
  • Furnace chamber with embedded ceramic muffle, high resistance to aggressive gasses and vapours
  • Dual shell housing made of textured stainless steel sheets with additional fan cooling for low surface temperature
  • Exclusive use of insulation materials without categorization according to EC Regulation No. 1272/2008 (CLP)
  • Optional flap door (L) which can be used as work platform or lift door (LT) with hot surface facing away from the operator
  • Adjustable working air inlet in the door
  • Exhaust air outlet in rear wall of furnace
  • Controller with touch operation B510 (5 programs with each 4 segments)
  • Solid state relays provide for lownoise operation
  • Defined application within the constraints of the operating instructions
  • NTLog Basic for Nabertherm controller: recording of process data with USB-flash drive

+ Technical Specifications

Features
  • Tmax 1100 °C
  • Muffle heated from four sides
  • Furnace chamber with embedded ceramic muffle, high resistance to aggressive gasses and vapours
  • Dual shell housing made of textured stainless steel sheets with additional fan cooling for low surface temperature
  • Exclusive use of insulation materials without categorization according to EC Regulation No. 1272/2008 (CLP)
  • Optional flap door (L) which can be used as work platform or lift door (LT) with hot surface facing away from the operator
  • Adjustable working air inlet in the door
  • Exhaust air outlet in rear wall of furnace
  • Controller with touch operation B510 (5 programs with each 4 segments)
  • Solid state relays provide for lownoise operation
  • Defined application within the constraints of the operating instructions
  • NTLog Basic for Nabertherm controller: recording of process data with USB-flash drive
Model Units L 9/11/SKM LT 9/11/SKM
Tmax in °C1 1100 1100
Inner dimensions in mm w 230 230
Inner dimensions in mm d 240 240
Inner dimensions in mm h 170 170
Volume in l 9 9
Outer dimensions2 in mm W 490 490
Outer dimensions2 in mm D 505 505
Outer dimensions2 in mm H 580 580+3203
Connected load5 in kW 3.7 3.7
Electrical connection 1-phase 1-phase
Weight in kg 50 50
Heating time in min4 75 75

1Recommended working temperature for processes with longer dwell times is 1000 °C
2External dimensions vary when furnace is equipped with additional equipment. Dimensions on request.
3Including opened lift door
4Heating time of the empty and closed furnace up to Tmax −100 K (connected to 230 V 1/N/PE)