Rolling Ball Tack Tester

Easy Determination Of Adhesive Tack

Easy Determination Of Adhesive Tack

Easy Determination Of Adhesive Tack

Model No. :

RBT-100

Model No. : RBT-100

Brand : Chem Instruments

This economical laboratory instrument precisely quantifies the quick-stick performance, or tack, of pressure-sensitive adhesives. It utilizes the inclined plane method to measure the distance a steel ball travels, providing consistent and reliable quality control data.

  • Solid aluminum construction ensures durability.

  • CNC machined inclined trough for accuracy.

  • Minimizes ball-to-tape bounce interference.

[acf_all_fields_accordion]

− Features

  • Solid aluminum construction, made to withstand many years of laboratory use
  • CNC machined inclined trough, engineered for a smooth and accurate test
  • Extra care taken in the design to minimize a “bounce” at the ball-to-tape interface
  • Simple loading

General Test Methods

  • Clean the inclined trough and ball with n-heptane or methyl alcohol.
  • Do not touch the trough or ball with fingers after cleaning.
  • Cut a 15″ long (380mm) by 2″ wide (50mm) specimen for testing purposes.
  • Arrange the test specimen at the end of the device, adhesive side up, in a straight line with the trough. The bottom of the device should hold downthe leading end of the specimen and a weight or other tape should hold the other end securely.
  • Ensure the device is level and aligned with the trough axis.
  • Use a lint free cloth to place the ball at the top of the inclined trough for testing.
  • Press the spring lever and allow the ball to roll down the trough.
  • Measure the distance between the end of the incline and the point where the ball stops.

+ Technical Specifications

+ Standards

+ Videos

Features
  • Solid aluminum construction, made to withstand many years of laboratory use
  • CNC machined inclined trough, engineered for a smooth and accurate test
  • Extra care taken in the design to minimize a “bounce” at the ball-to-tape interface
  • Simple loading

General Test Methods

  • Clean the inclined trough and ball with n-heptane or methyl alcohol.
  • Do not touch the trough or ball with fingers after cleaning.
  • Cut a 15″ long (380mm) by 2″ wide (50mm) specimen for testing purposes.
  • Arrange the test specimen at the end of the device, adhesive side up, in a straight line with the trough. The bottom of the device should hold downthe leading end of the specimen and a weight or other tape should hold the other end securely.
  • Ensure the device is level and aligned with the trough axis.
  • Use a lint free cloth to place the ball at the top of the inclined trough for testing.
  • Press the spring lever and allow the ball to roll down the trough.
  • Measure the distance between the end of the incline and the point where the ball stops.
Height 2.5″ (6.5 cm)
Width 9″ (23 cm)
Depth 2.5″ (6.5 cm)
Weight 1 lbs (0.5 Kg)