TCT-1 | PL 2

Temperature Control For Powder Coating Gel Time

Temperature Control For Powder Coating Gel Time

Temperature Control For Powder Coating Gel Time

Model No. :

Brand : Gel Instrument

The TCT-1 | PL 2 is a high-precision thermal control system specifically engineered for determining the reactivity and polymerization time of thermosetting resins. This model features an integrated heating plate and a sophisticated digital controller to maintain exact temperature stability during critical B-time and gelation analysis.

  • Precision heating for resin reactivity testing

  • Integrated digital stopwatch for process timing

  • Accurate temperature regulation via Pt100 sensor

[acf_all_fields_accordion]

− Features

  • Dual-Component Design: Consists of a high-performance heating plate assembly paired with a separate digital control unit, ensuring heat isolation for the electronic components and long-term durability.
  • Superior Temperature Stability: Employs an internal Pt100 sensor and an external Type K thermocouple to monitor surface heat, maintaining an operating range up to 200°C with high precision.
  • Surface Configuration: The standard PL 2 heating plate is designed with specialized depressions to accommodate resin samples, facilitating uniform heat transfer and consistent testing conditions according to ISO standards.
  • Integrated Chronometer: Features a built-in digital timer on the controller front panel, allowing operators to trigger the stopwatch simultaneously with sample application for exact B-time determination.
  • Modular Connectivity: Designed to interface with advanced data logging systems, providing a scalable solution for laboratory R&D and industrial quality control in resin manufacturing.

+ Technical Specifications

+ Standards

Features
  • Dual-Component Design: Consists of a high-performance heating plate assembly paired with a separate digital control unit, ensuring heat isolation for the electronic components and long-term durability.
  • Superior Temperature Stability: Employs an internal Pt100 sensor and an external Type K thermocouple to monitor surface heat, maintaining an operating range up to 200°C with high precision.
  • Surface Configuration: The standard PL 2 heating plate is designed with specialized depressions to accommodate resin samples, facilitating uniform heat transfer and consistent testing conditions according to ISO standards.
  • Integrated Chronometer: Features a built-in digital timer on the controller front panel, allowing operators to trigger the stopwatch simultaneously with sample application for exact B-time determination.
  • Modular Connectivity: Designed to interface with advanced data logging systems, providing a scalable solution for laboratory R&D and industrial quality control in resin manufacturing.

Heating Block PL2

Heating Block PL2 according to DIN ISO 8130-6 Art-No. 200.30.50
Temperature range 30.0°C … 200.0°C (on request 250.0°C)
Power supply 230V / 50 Hz or 115V / 60 Hz, ± 10%
Heat power 400W
Dimensions Diameter 130mm, height 85mm
Weight an about ~ 1.66 kg
Melting salts Various e.g. 140 – 180°C

 

Temperature control unit TCT-1 with stopwatch

TCT-1 Item No. 200.41.01
Power supply 230VAC, 50/60 Hz or 115VAC 50 / 60 Hz
Power consumption 600W with connected heating block PL2
Connectors 8 Pole Binder connector for connecting the heating block PL2
Sensor Pt100, EN60751
Temperature range 30.0°C … 200.0°C (on request 250.0°C)
Accuracy ± 0.5°C
Additional measurement 1 x type K, EN 60584, for the surface temperature
Main switch on the rear panel
Fuses 2 pieces F3,15A L250VAC
Dimensions 257 x 271 x 103 mm
Weight ~ 3.4 kg
Electrical safety 2006/95/EC, 2004/108/EC, EN 61010-1:2001, 61010-2-010:2003 EN, EN 61326-1:2006

 

Integrated Stopwatch

Range 0 … 99H: 59M: 59S
Accuracy ± 1 second / 2 hours